Z-0935-04
Synchron Systems Drug Calibrator 1 (Carbamaxepine, Phenobarbitol, Phentoin, Theophylline, and Valproic Acid)
Siemens Healthcare Diagnostics Inc. / initiated 2003-06-11 / Terminated
Beckman Coulter Inc began a recall of Synchron Systems Drug Calibrator 1 (Carbamaxepine, Phenobarbitol, Phentoin, Theophylline, and Valproic Acid) on . The reason FDA records is "Product was found to produce higher recoveries by about 7-12% compared to other lots". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0935-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Synchron Systems Drug Calibrator 1 (Carbamaxepine, Phenobarbitol, Phentoin, Theophylline, and Valproic Acid).
- Product code
- KLT Enzyme Immunoassay, Carbamazepine
- Reason for recall
- Product was found to produce higher recoveries by about 7-12% compared to other lots.
- FDA root cause
- Other
- Action
- Users were requested to discard and ask for replacement of the lot in question by letter.
- Quantity in commerce
- unknown
- Distribution
- Nationwide and Canada
- Product codes and lots
- Lot M207140