Z-0929-04

Concha IV Plus Heated Humidifier

Hudson Respiratory Care, Inc. / initiated 2004-04-20 / Terminated / root cause: software

Hudson Respiratory Care Inc began a recall of Concha IV Plus Heated Humidifier on . The reason FDA records is "Software malfunction. The humidifier did not retain the clinician selected default settings for temperature gradient, after the unit had been placed in the "pause" mode". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0929-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hudson Respiratory Care, Inc.
Product
Concha IV Plus Heated Humidifier
Product code
BTT Humidifier, Respiratory Gas, (Direct Patient Interface)
Reason for recall
Software malfunction. The humidifier did not retain the clinician selected default settings for temperature gradient, after the unit had been placed in the "pause" mode.
Root cause tag
software
FDA root cause
Other
Action
Firm will call hospitals to arrange for a visit to update the software.
Quantity in commerce
1049
Distribution
Nationwide.
Product codes and lots
All products manufactured with software 11738 -rev.03, Version V13.6, 252 A. This represents all units currently in the field.