Z-0929-04
Concha IV Plus Heated Humidifier
Hudson Respiratory Care, Inc. / initiated 2004-04-20 / Terminated / root cause: software
Hudson Respiratory Care Inc began a recall of Concha IV Plus Heated Humidifier on . The reason FDA records is "Software malfunction. The humidifier did not retain the clinician selected default settings for temperature gradient, after the unit had been placed in the "pause" mode". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0929-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hudson Respiratory Care, Inc.
- Product
- Concha IV Plus Heated Humidifier
- Product code
- BTT Humidifier, Respiratory Gas, (Direct Patient Interface)
- Reason for recall
- Software malfunction. The humidifier did not retain the clinician selected default settings for temperature gradient, after the unit had been placed in the "pause" mode.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Firm will call hospitals to arrange for a visit to update the software.
- Quantity in commerce
- 1049
- Distribution
- Nationwide.
- Product codes and lots
- All products manufactured with software 11738 -rev.03, Version V13.6, 252 A. This represents all units currently in the field.