Z-1054-04

Clintec Micromix Compounder, an I.V. solution compounder; Baxter Healthcare Corporation, Clintec Nutrition...

Baxter Healthcare Corporation / initiated 2004-04-26 / Terminated

Baxter Healthcare Corp. began a recall of Clintec Micromix Compounder, an I.V. solution compounder; Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield, IL 60015 USA; product code... on . The reason FDA records is "The Micromix compounder may have ingredient delivery outside of the stated accuracy limits if the Vlier pin is not properly adjusted". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1054-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Clintec Micromix Compounder, an I.V. solution compounder; Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield, IL 60015 USA; product code 2M8290 - 1.0 mL accuracy, and product code 2M8290L - less than 1.0 mL accuracy
Product code
NEP--
Reason for recall
The Micromix compounder may have ingredient delivery outside of the stated accuracy limits if the Vlier pin is not properly adjusted
FDA root cause
Other
Action
Baxter began telephoning their customers on 4/26/04, advising them that an improperly adjusted Vlier pin could result in an ingredient delivery outside of stated accuracy limits. The accounts were requested to immediately stop using the Micromix Compounder and return the device to Baxter, and to evaluate all TPN formualtions that have been compounded but not administered to determine if any ingredients were delivered at a programmed volume of less than 1 mL. The account was advised to discard those solutions and recompound them using manual techniques. Follow-up recall letters were faxed to the accounts on the same date, reiterating the telephone calls. Baxter sent second recall letters dated 2/16/05 to the 15 customers who had not yet returned their Micromix Compounder as of 2/16/05, along with a copy of the 4/26/04 recall letter. The customers were informed of the new issue with the Micromix Compounder not recognizing decimal points for delivery volumes below 1 mL, and were again requested to return the Micromix Compounder and the handheld keypad to Baxter so that they may be corrected when the Micromix is reintroduced the market.
Quantity in commerce
500 devices
Distribution
Nationwide and internationally through Baxter subsidiaries to the United Kingdom, Germany, Chile, Canada, Brazil, El Salvador, Taiwan and Korea.
Product codes and lots
product codes 2M8290 and 2M8290L, all serial numbers The recall extension concerns those units that had not been returned as of 2/16/05: serial numbes 10025, 11163, 11263, 11298, 11254, 11280, 10075, 11279, 11265, 11264, 10065, 11251, 11065, T12054, 10051