Z-0944-04

Deikos A+

BIOTRONIK AG / initiated 2004-04-29 / Terminated

Biotronik Inc began a recall of Deikos A+ on . The reason FDA records is "Potential for device to not deliver full amount of energy". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0944-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BIOTRONIK AG
Product
Deikos A+
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Potential for device to not deliver full amount of energy
FDA root cause
Other
Action
On 4/29/04 the firm issued 'Dear Doctor' letters to the physicians. The letter instructs the physician to schedule an immediate follow-up with all affected patients to determine if their devices are being impacted by the passivation buildup and to determine if device replacement is indicated.
Quantity in commerce
6
Distribution
Devices were distributed to physicians and medical centers throughout the U.S. The firm does not distribute internationally.
Product codes and lots
Model Number 334 341 Serial numbers: 78062958, 78062960, 78062961, 78062962, 78062997, 78063002