Z-0996-04
Tecan clinical workstation; Tecan Genesis 150 front end, Roche catalog number 03585379001
Roche Diagnostics GmbH / initiated 2004-03-05 / Terminated
Roche Diagnostics Corp. began a recall of Tecan clinical workstation; Tecan Genesis 150 front end, Roche catalog number 03585379001 on . The reason FDA records is "If the files are not periodically purged from the TCW computer, it may result in patient identification/sample results mismatches". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0996-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Tecan clinical workstation; Tecan Genesis 150 front end, Roche catalog number 03585379001.
- Product code
- JQW Station, Pipetting And Diluting, For Clinical Use
- Reason for recall
- If the files are not periodically purged from the TCW computer, it may result in patient identification/sample results mismatches.
- FDA root cause
- Other
- Action
- Roche sent each of their consignees revised Tecan database maintenance instruction via bulletin 04-042 dated 3/5/04.
- Quantity in commerce
- 14
- Distribution
- Nationwide. California, Connecticut, Georgia, Illinois, Indiana, Maryland, Massachusetts, Michigan, New Jersey, North Carolina, Ohio, Pennsylvania, Tennessee, Texas and Wisconsin.
- Product codes and lots
- Serial numbers 5524, 6177, 6224, 6376, 6393, 6395, 6397, 6557, 7150, 7349, 7350, 7594, 7616 and 7724.