Z-1015-04

Dinalink brand Communication Cable which is an accessory to a medical device Dinalike which is an isolated...

GE Medical Systems Information Technologies, Inc. / initiated 2004-01-27 / Terminated

GE Medical Systems Information Technologies began a recall of Dinalink brand Communication Cable which is an accessory to a medical device Dinalike which is an isolated protocol converter used to concert the serial data... on . The reason FDA records is "The potential exists for external electrical noise to be processed as irregular pacing thereby masking potential asystole events". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1015-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Medical Systems Information Technologies, Inc.
Product
Dinalink brand Communication Cable which is an accessory to a medical device Dinalike which is an isolated protocol converter used to concert the serial data from a DINAMAP Monitor to that of the APEXPRO telemetry transmitter.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
The potential exists for external electrical noise to be processed as irregular pacing thereby masking potential asystole events.
FDA root cause
Other
Action
The firm mailed an 'Urgent Medical Device Correction' letter to consignees on January 27, 2004 advising them of the external electrical noise problem and informing them to not use the Dinalink ILC1931 with patients with pace makers until a fix or replacement can be provided. By 5/17/2004 corrective replacement cables had been provided to all consignees.
Quantity in commerce
36
Distribution
Product was distributed to 7 domestic hospitals in PA, SC, GA, FL, WI, SD AND WY.
Product codes and lots
No serial numbers but the Dinalink protocol converter and cable has the product identifier of ''ILC1931''