Z-0990-04
Power Heart AED G3 Automated External Defibrillator IntelliSense (Lithium) Battery, Model 9142
ZOLL Medical Corporation / initiated 2004-05-04 / Terminated
Cardiac Science, Inc. began a recall of Power Heart AED G3 Automated External Defibrillator IntelliSense (Lithium) Battery, Model 9142 on . The reason FDA records is "Some of the recalled defibrillator batteries contain an incorrect fuse which could open resulting in a defibrillator which will not work". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0990-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ZOLL Medical Corporation
- Product
- Power Heart AED G3 Automated External Defibrillator IntelliSense (Lithium) Battery, Model 9142
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- Some of the recalled defibrillator batteries contain an incorrect fuse which could open resulting in a defibrillator which will not work.
- FDA root cause
- Other
- Action
- A recall letter, dated May 10, 2004, stated that a replacement battery is enclosed with the letter, and instructed consignees to install the new battery in their defibrilllator and return the old battery.
- Distribution
- The batteries were distributed with the defibrillators nationwide in the United States and to foreign consignees in Europe, Asia, Latin America, the Middle East and Africa.
- Product codes and lots
- The recalled Model 9142 Batteries have the following lot numbers: 7848, 7849, 7850, 7853, 7854, 7855, 7857, 7858, 7859, 7860, 7861, 7863, 7866, 7867, 7868, 7870, 7871, 7872, and 7874.