Z-1025-04

Ascension PIP instrument tray base, manufactured by Ascension Orthopedics, Austin, Texas

Ascension Orthopedics, Inc. / initiated 2003-10-02 / Terminated / root cause: packaging

Ascension Orthopedics, Inc began a recall of Ascension PIP instrument tray base, manufactured by Ascension Orthopedics, Austin, Texas on . The reason FDA records is "Cracks in the tray due to excessive use can cause perforations in the outer wrap used to maintain instrument tray sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1025-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ascension Orthopedics, Inc.
Product
Ascension PIP instrument tray base, manufactured by Ascension Orthopedics, Austin, Texas
Product code
NEG Finger Semi-Constrained Pyrolytic Carbon Uncemented Prosthesis
Reason for recall
Cracks in the tray due to excessive use can cause perforations in the outer wrap used to maintain instrument tray sterility.
Root cause tag
packaging
FDA root cause
Other
Action
Recall was initiated via letter included with replacement trays begining in October 2003.
Quantity in commerce
107 trays
Distribution
Nationwide distrbution to IA, TX, OH, FL, IN, MN, CO, WA, CA, MO, and to a consignee in Ontario, Canada