Z-1077-04
GemStar I.V. Administration Pump Set with Orange Polyethylene-Lined Light Resistant Tubing, List Number: 13271
Abbott Labs / initiated 2004-05-11 / Terminated
Abbott Labs began a recall of GemStar I.V. Administration Pump Set with Orange Polyethylene-Lined Light Resistant Tubing, List Number: 13271 on . The reason FDA records is "The firm's internal testing that revealed the potential air-in-line may go undetected when the set is used at temperatures above 25C (77 F)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1077-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Labs
- Product
- GemStar I.V. Administration Pump Set with Orange Polyethylene-Lined Light Resistant Tubing, List Number: 13271
- Product code
- FRN Pump, Infusion
- Reason for recall
- The firm's internal testing that revealed the potential air-in-line may go undetected when the set is used at temperatures above 25C (77 F).
- FDA root cause
- Other
- Action
- Recall initiated May 11, 2004. Letters were sent to consignees and wholesalers.
- Quantity in commerce
- 18,432 units
- Distribution
- Total number of consignees that received product: 13. Five are wholesale dealers, None are distributors, 8 are retailers. Product has been distributed worldwide, and in the United States: AR, Washington DC, FL, IL, KY, LA, MA, MD, MN, MO, NJ, TX. Also sent to National Institute of Health, Bethesda, MD, (301) 496-4000
- Product codes and lots
- 95-156-5H, 05-027-5H, 09-208-5H