Synchron Liquid Comprehensive Chemistry Control Serum. Part numbers: 660710, 657365

Siemens Healthcare Diagnostics Inc. / initiated 2003-04-01 / Terminated

Beckman Coulter Inc began a recall of Synchron Liquid Comprehensive Chemistry Control Serum. Part numbers: 660710, 657365 on . The reason FDA records is "The quality control samples for Vancomycin would not recover correctly when compared to the S.I. unit values listed in the incorrectly printed product insert. The incorrectly printed values were higher than they should...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1151-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Synchron Liquid Comprehensive Chemistry Control Serum. Part numbers: 660710, 657365.
Product code
LEH Radioimmunoassay, Vancomycin
Reason for recall
The quality control samples for Vancomycin would not recover correctly when compared to the S.I. unit values listed in the incorrectly printed product insert. The incorrectly printed values were higher than they should be by a factor of 3 (three).
FDA root cause
Other
Quantity in commerce
9730
Product codes and lots
Lots: M202040, M206290