Z-1151-04
Synchron Liquid Comprehensive Chemistry Control Serum. Part numbers: 660710, 657365
Siemens Healthcare Diagnostics Inc. / initiated 2003-04-01 / Terminated
Beckman Coulter Inc began a recall of Synchron Liquid Comprehensive Chemistry Control Serum. Part numbers: 660710, 657365 on . The reason FDA records is "The quality control samples for Vancomycin would not recover correctly when compared to the S.I. unit values listed in the incorrectly printed product insert. The incorrectly printed values were higher than they should...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1151-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Synchron Liquid Comprehensive Chemistry Control Serum. Part numbers: 660710, 657365.
- Product code
- LEH Radioimmunoassay, Vancomycin
- Reason for recall
- The quality control samples for Vancomycin would not recover correctly when compared to the S.I. unit values listed in the incorrectly printed product insert. The incorrectly printed values were higher than they should be by a factor of 3 (three).
- FDA root cause
- Other
- Quantity in commerce
- 9730
- Product codes and lots
- Lots: M202040, M206290