Vigil Lipid Control, Levels 1,2,3 and 4 for ApoB Reagent

Siemens Healthcare Diagnostics Inc. / initiated 2003-08-18 / Terminated / root cause: labeling and UDI

Beckman Coulter Inc began a recall of Vigil Lipid Control, Levels 1,2,3 and 4 for ApoB Reagent on . The reason FDA records is "Values for ApoB were not properly assigned on the package insert. Customers were not able to recover the analyte within the stated values". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1156-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Vigil Lipid Control, Levels 1,2,3 and 4 for ApoB Reagent.
Product code
MSS Folders And Injectors, Intraocular Lens (Iol)
Reason for recall
Values for ApoB were not properly assigned on the package insert. Customers were not able to recover the analyte within the stated values.
Root cause tag
labeling and UDI
FDA root cause
Other
Quantity in commerce
5275
Product codes and lots
CX: Level 1 -M206191, Level 2-M206192, Level 3-M206193, Level 4-M206194. LX: Level 1-M301121, Level 2- M301122, Level 3- M301122, Level 4-M301124.