Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0977-04

Mentor Smooth Round Becker 50 Expander, manufactured by Mentor, Irving, Texas

Mentor / initiated 2003-12-16 / Terminated

Mentor Texas, Inc began a recall of Mentor Smooth Round Becker 50 Expander, manufactured by Mentor, Irving, Texas on . The reason FDA records is "Device was distributed without the dome part accessory, part number 350-DOMPK". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0977-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Mentor
Product
Mentor Smooth Round Becker 50 Expander, manufactured by Mentor, Irving, Texas
Product code
FTN Plastic Surgery And Accessories Kit
Reason for recall
Device was distributed without the dome part accessory, part number 350-DOMPK
FDA root cause
Other
Action
The recall letters for Lot 269402 were sent on 12/16/2003. The recall letters for Lot 264358 were sent on 02/06/2004. Each letter included 5 dome pack accessories as a temporary corrective action.
Quantity in commerce
51 units
Distribution
Nationwide. Recalled product was shipped domestically to accounts in AZ, AR, CA, GA, IL, IN, KY, MD, MI, OH, OK, OR, TX, and NY. There was no foreign distribution.
Product codes and lots
Lot numbers 269402 and 264358