Z-1032-04
Infant Ventilators with Electronic Manometer. Models IV-100B; IV-200 and IV-200 SAVI
Sechrist Industries Inc / initiated 2004-05-20 / Terminated
Sechrist Industries Inc began a recall of Infant Ventilators with Electronic Manometer. Models IV-100B; IV-200 and IV-200 SAVI on . The reason FDA records is "Electronic manometer malfunctions. Manometer may be illegible. The alarm for power failure may function only briefly, or not at all". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1032-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sechrist Industries Inc
- Product
- Infant Ventilators with Electronic Manometer. Models IV-100B; IV-200 and IV-200 SAVI.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Electronic manometer malfunctions. Manometer may be illegible. The alarm for power failure may function only briefly, or not at all.
- FDA root cause
- Other
- Action
- Recall notification contains a test method to be used to identify defective units. Return notification that ventilator is defective will result in replacement of the electronic manometer.
- Quantity in commerce
- 240
- Distribution
- Nationwide and to Malaysia, Japan, Kuwait, Sri Lanka, South Africa, Guatemala, Mexico, Morocco, Indonesia, Korea, Ecuador, Colombia, Portugal, Oman, Switzerland, Spain, Zimbabwe, Belgium, Brazil, Thailand, Argentina, Costa Rica, Taiwan, ROC, India, Russia, Saudi Arabia, Pakistan, Caribbean, Venezuela.
- Product codes and lots
- Restricted to those ventilators with production serial numbers beginning with a ''00'' prefix.