Z-1071-04

NxStage Cartridge Express. A disposable extracorporeal blood and fluid pathway and a hemo filter. Model #...

NxStage Medical Inc / initiated 2004-04-14 / Terminated

NxStage Medical, Inc. began a recall of NxStage Cartridge Express. A disposable extracorporeal blood and fluid pathway and a hemo filter. Model # CAR-154 Cartridge Express, 6 units per case on . The reason FDA records is "Pinhole leaks on the replacement (RF) side FMP bag in certain lots of its disposable cartridge balance chambers resulting in loss of dialysate or ultrafiltrate". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1071-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
NxStage Medical Inc
Product
NxStage Cartridge Express. A disposable extracorporeal blood and fluid pathway and a hemo filter. Model # CAR-154 Cartridge Express, 6 units per case.
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Pinhole leaks on the replacement (RF) side FMP bag in certain lots of its disposable cartridge balance chambers resulting in loss of dialysate or ultrafiltrate
FDA root cause
Other
Action
Voluntary recall notice dated 4/14/04 sent UPS, overnight with signature required to both ICU/CCU (Acute accounts) and Chronic accounts. Firm to remove and replace product that contains a disposable component from a new supplier that has demonstrated to be less reliable resulting in pinhole leaks.
Quantity in commerce
552 units
Distribution
CA, IL, IN, MD, IA, MO, NJ, OH, VT
Product codes and lots
Lot Numbers from and including FG-031216-01 to FG-040322-01