Cook brand Zenith AAA Converter with the H&L-B One-Shot Introducer System, preloaded converter - proximal...

Cook Incorporated / initiated 2004-06-10 / Terminated

Cook, Inc. began a recall of Cook brand Zenith AAA Converter with the H&L-B One-Shot Introducer System, preloaded converter - proximal diameter 24 mm - distal diameter 12 mm - length 80... on . The reason FDA records is "Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1160-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Cook brand Zenith AAA Converter with the H&L-B One-Shot Introducer System, preloaded converter - proximal diameter 24 mm - distal diameter 12 mm - length 80 mm; Reorder no. ESC-24-12-80.
Product code
MIH System, Endovascular Graft, Aortic Aneurysm Treatment
Reason for recall
Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use.
FDA root cause
Other
Action
The firm issued recall letters to customers on 6/10/04. The recall was extended and URGENT EXTENSION OF PRODUCT RECALL notices were sent out on June 23, 2004.
Quantity in commerce
7
Distribution
Nationwide, Canada, Denmark and Japan.
Product codes and lots
Lots 1356524, 1356532, 1356519, 1356525, 1356523, 1356522 and 1356529.