Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1073-04

TefGen is an implant material which is intended to be used as a temporary space-making barrier over bone or...

Lifecore Biomedical LLC / initiated 2004-06-03 / Terminated / root cause: labeling and UDI

Lifecore Biomedical, Inc. began a recall of TefGen is an implant material which is intended to be used as a temporary space-making barrier over bone or other tissue on . The reason FDA records is "Packaged product labeled catalog #TFS30, lot number 002419 actually contained catalog #TFR60, lot number 002441and vice versa". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1073-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Lifecore Biomedical LLC
Product
TefGen is an implant material which is intended to be used as a temporary space-making barrier over bone or other tissue.
Product code
LYC Bone Grafting Material, Synthetic
Reason for recall
Packaged product labeled catalog #TFS30, lot number 002419 actually contained catalog #TFR60, lot number 002441and vice versa.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Each facility was notified of the recall by letter on 06/03/2004.
Quantity in commerce
200 of lot #002419 and 55 of lot #002441
Distribution
Throughout the U.S., Germany, France, Korea, Columbia, Chile and Korea
Product codes and lots
catalog #TFS30 lot number 002419 and catalog #TFR60 lot number 002441