Dispersive Electrodes packaged in an 8x10 plastic container

Boston Scientific Corporation / initiated 2004-06-16 / Terminated

Rita Medical Systems began a recall of Dispersive Electrodes packaged in an 8x10 plastic container on . The reason FDA records is "Inaccurate temperature readings may be displayed by the device, actual temperature could be 2.5-3 degrees C higher than displayed". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1076-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Dispersive Electrodes packaged in an 8x10 plastic container.
Product code
JOS Electrode, Electrosurgical
Reason for recall
Inaccurate temperature readings may be displayed by the device, actual temperature could be 2.5-3 degrees C higher than displayed.
FDA root cause
Other
Action
Each hospital that has been shipped the devices will receive a recall letter during the week of june 21, 2004 with materials for return shipping. Field representatives will receive notice of the recall, and will return any unused devices and document consumed devices.
Quantity in commerce
468 sets (each set consists of two pads)
Distribution
Product was shipped to 18 field representatives and 87 customers (hospitals and medical centers
Product codes and lots
Lot codes affected: Y110503-1 and Y103003-2. 510k number : k031926 Contract manufacturer, Katecho, Inc, registration number 1932007