Z-1233-04
3 Series Full Body Phototherapy Device with Smart Touch Control System, Models #PASNBSM-2424
The Daavlin Distributing Co. / initiated 2003-09-01 / Terminated / root cause: software
Daavlin Distributing Company began a recall of 3 Series Full Body Phototherapy Device with Smart Touch Control System, Models #PASNBSM-2424 on . The reason FDA records is "Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1233-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- The Daavlin Distributing Co.
- Product
- 3 Series Full Body Phototherapy Device with Smart Touch Control System, Models #PASNBSM-2424.
- Product code
- FTC Light, Ultraviolet, Dermatological
- Reason for recall
- Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- The firm has software patches to correct the programming problems as they are reported by the users. These patches are being installed either by a service representative for the manufacturer or by the customer
- Quantity in commerce
- 2 units distributed domestically.
- Distribution
- Nation wide plus Canada, the Philippines, Saudi Arabia, Turkey & Germany
- Product codes and lots
- All Touch Control System, Model #PASNBSM-2424 units manufactured between 6/5/03 and 4/30/04.