Z-1060-04
ACCU-CHEK Inform meter; catalog No. 3035123
Roche Diagnostics GmbH / initiated 2004-06-30 / Terminated
Roche Diagnostics Corp. began a recall of ACCU-CHEK Inform meter; catalog No. 3035123 on . The reason FDA records is "Exposure to excessive disinfectant or cleaning solution may cause overheating, melting, or smoke". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1060-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- ACCU-CHEK Inform meter; catalog No. 3035123.
- Product code
- NBW System, Test, Blood Glucose, Over The Counter
- Reason for recall
- Exposure to excessive disinfectant or cleaning solution may cause overheating, melting, or smoke.
- FDA root cause
- Other
- Action
- A recall notice dated 6/30/04 was sent to each customer instructing them on proper cleaning methods, asking them to inspect their units for damage, and providing warning stickers for attachment to the base units.
- Distribution
- Nationwide.
- Product codes and lots
- All units.