Z-1064-04

Osmetech OPTI-CCA Blood Gas Analyzer, Single Use, BP7564 OPTIE-Glu Cassettes.The cassettes are labeled, Roche...

Opti Medical Systems Inc / initiated 2004-06-22 / Terminated

Osmetech Inc began a recall of Osmetech OPTI-CCA Blood Gas Analyzer, Single Use, BP7564 OPTIE-Glu Cassettes.The cassettes are labeled, Roche OPTICCA Cassettes E-Glu. There are 25 cassettes... on . The reason FDA records is "In some cases patient samples with high glucose concentrations, above the reportable range of the OPTI-CCA Analyzer, may be erroneously reported as 'Low'". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1064-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Opti Medical Systems Inc
Product
Osmetech OPTI-CCA Blood Gas Analyzer, Single Use, BP7564 OPTIE-Glu Cassettes.The cassettes are labeled, Roche OPTICCA Cassettes E-Glu. There are 25 cassettes in each box.
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
In some cases patient samples with high glucose concentrations, above the reportable range of the OPTI-CCA Analyzer, may be erroneously reported as 'Low'.
FDA root cause
Other
Action
Consignees were notified by letter on 6/22/04. A separate Recall Notice was mailed to distributors outside the United States with instructions on issuing the recall notice to their affected customers in the appropriate language.
Quantity in commerce
1087 boxes
Distribution
Nationwide and Argentina, Austria, Bolivia, France, Greece, Germany, Korea, India, Italy, Jordan, Netherlands, Sweden, Switzerland.
Product codes and lots
All OPTI-CCA Analyzers with software version 2.20 through 2.24 are affected. Catalog Number BP7564, Lot #351601, 401600, 401601, 405600, 405601, 406600, 406601, 405600, 405601, 410600, 412600, 415600, 415601, 417600, 419600, 419601, 419602 and 419603.