Z-1064-04
Osmetech OPTI-CCA Blood Gas Analyzer, Single Use, BP7564 OPTIE-Glu Cassettes.The cassettes are labeled, Roche...
Opti Medical Systems Inc / initiated 2004-06-22 / Terminated
Osmetech Inc began a recall of Osmetech OPTI-CCA Blood Gas Analyzer, Single Use, BP7564 OPTIE-Glu Cassettes.The cassettes are labeled, Roche OPTICCA Cassettes E-Glu. There are 25 cassettes... on . The reason FDA records is "In some cases patient samples with high glucose concentrations, above the reportable range of the OPTI-CCA Analyzer, may be erroneously reported as 'Low'". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1064-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Opti Medical Systems Inc
- Product
- Osmetech OPTI-CCA Blood Gas Analyzer, Single Use, BP7564 OPTIE-Glu Cassettes.The cassettes are labeled, Roche OPTICCA Cassettes E-Glu. There are 25 cassettes in each box.
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- In some cases patient samples with high glucose concentrations, above the reportable range of the OPTI-CCA Analyzer, may be erroneously reported as 'Low'.
- FDA root cause
- Other
- Action
- Consignees were notified by letter on 6/22/04. A separate Recall Notice was mailed to distributors outside the United States with instructions on issuing the recall notice to their affected customers in the appropriate language.
- Quantity in commerce
- 1087 boxes
- Distribution
- Nationwide and Argentina, Austria, Bolivia, France, Greece, Germany, Korea, India, Italy, Jordan, Netherlands, Sweden, Switzerland.
- Product codes and lots
- All OPTI-CCA Analyzers with software version 2.20 through 2.24 are affected. Catalog Number BP7564, Lot #351601, 401600, 401601, 405600, 405601, 406600, 406601, 405600, 405601, 410600, 412600, 415600, 415601, 417600, 419600, 419601, 419602 and 419603.