Z-1082-04

Pipette Sample Tip, Product part #65910, packaged in cardboard cartons containing 4 plastic racks of 108...

Dynex Technologies, Inc. / initiated 2004-06-11 / Terminated

Dynex Technologies, Inc. began a recall of Pipette Sample Tip, Product part #65910, packaged in cardboard cartons containing 4 plastic racks of 108 pipette tips per rack. Pipette tips are accessory... on . The reason FDA records is "An accessory sampling tip for an in vitro diagnostic test instrument may be inadequately sealed and may cause false assay patient results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1082-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Dynex Technologies, Inc.
Product
Pipette Sample Tip, Product part #65910, packaged in cardboard cartons containing 4 plastic racks of 108 pipette tips per rack. Pipette tips are accessory component for Dynex Technologies proprietary automated DSX system.
Product code
JQW Station, Pipetting And Diluting, For Clinical Use
Reason for recall
An accessory sampling tip for an in vitro diagnostic test instrument may be inadequately sealed and may cause false assay patient results.
FDA root cause
Other
Action
The recalling firm notified consignees by faxed letter on 6/11/04 and follow-up letter by mail 6/24/04. The notice advised of vacuum failure associated with lot 412410 of sample tips and recommended quarantine of the defective tips. Consignees were requested to return response form. The 6/24/04 notification recommends that consignees discard/destroy lot 412410 and asks distributors to notify their customers. The firm is working with supplier of the pipette tips to replace recalled product.
Quantity in commerce
1749 cartons
Distribution
Pipette tips were sold to seven OEMs for distribution to end users in the US and internationally.
Product codes and lots
Lot 412410 printed on carton