Z-1232-04

Boehringer Replacement Gauge Model 9071, Lot Number 200405

Boehringer Laboratories, LLC / initiated 2004-05-19 / Terminated / root cause: manufacturing process

Boehringer Laboratories Inc began a recall of Boehringer Replacement Gauge Model 9071, Lot Number 200405 on . The reason FDA records is "The suction regulator incorporates a gauge for the verification of suction levels. The particular lot of gauges used in the production of the regulators exhibits a potential to develop a leak and cause the delivered...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1232-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boehringer Laboratories, LLC
Product
Boehringer Replacement Gauge Model 9071, Lot Number 200405.
Product code
GCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered
Reason for recall
The suction regulator incorporates a gauge for the verification of suction levels. The particular lot of gauges used in the production of the regulators exhibits a potential to develop a leak and cause the delivered suction to be inaccurate.
Root cause tag
manufacturing process
FDA root cause
Other
Action
All direct accounts were contacted by the recalling firm''s Customer Service Representative to determine the location of the affected product. The CSR arrange for replacement proudct.
Quantity in commerce
9 units
Distribution
The recalled product was distributed to the following states: AL, FL, MA, MI, OR, PA, TN,
Product codes and lots
Lot Number 200405