Z-1446-04
AXIOM Sensis Report Workstation. Programmable diagnostic computer. Model number 6627637
Siemens Medical Solutions USA, Inc. / initiated 2004-05-17 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of AXIOM Sensis Report Workstation. Programmable diagnostic computer. Model number 6627637 on . The reason FDA records is "Software issue. Allows for users to enter date in Pre-Cath Holding Area prior to registering the patient in the Cath Lab". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1446-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- AXIOM Sensis Report Workstation. Programmable diagnostic computer. Model number 6627637
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- Software issue. Allows for users to enter date in Pre-Cath Holding Area prior to registering the patient in the Cath Lab.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- The recalling firm issued a Customer Safety Advisory to their customers via certified mail on 5/17/04 to inform them of the problem. The advisory provides Update Instructions AX031/04/S to prevent this problem from occurring until the software upgrade is installed to correct the problem.
- Quantity in commerce
- 234 units
- Distribution
- The product was shipped to 43 medical facilities nationwide and one government account in KY.
- Product codes and lots
- Serial Numbers 1002, 1034, 1100-1253, 1300-1504