Z-1161-04

Cook brand Zenith AAA Endovascular Graft Iliac Leg Extensions - diameter 14 mm - length 55 mm -introducer...

Cook Incorporated / initiated 2004-06-10 / Terminated

Cook, Inc. began a recall of Cook brand Zenith AAA Endovascular Graft Iliac Leg Extensions - diameter 14 mm - length 55 mm -introducer sheath size 14 FR; Reorder no. ESLE-14-55-TAG on . The reason FDA records is "Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1161-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Cook brand Zenith AAA Endovascular Graft Iliac Leg Extensions - diameter 14 mm - length 55 mm -introducer sheath size 14 FR; Reorder no. ESLE-14-55-TAG.
Product code
MIH System, Endovascular Graft, Aortic Aneurysm Treatment
Reason for recall
Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use.
FDA root cause
Other
Action
The firm issued recall letters to customers on 6/10/04. The recall was extended and URGENT EXTENSION OF PRODUCT RECALL notices were sent out on June 23, 2004.
Quantity in commerce
1
Distribution
Nationwide, Canada, Denmark and Japan.
Product codes and lots
Lot F1575173.