Z-1169-04
Cook brand Zenith AAA Bifurcated main body graft with the H&L-B One-Shot Introducer System, preloaded...
Cook Incorporated / initiated 2004-06-10 / Terminated
Cook, Inc. began a recall of Cook brand Zenith AAA Bifurcated main body graft with the H&L-B One-Shot Introducer System, preloaded bifurcated main body graft - aortic graft diameter 24 mm... on . The reason FDA records is "Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1169-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Cook brand Zenith AAA Bifurcated main body graft with the H&L-B One-Shot Introducer System, preloaded bifurcated main body graft - aortic graft diameter 24 mm - iliac leg diameters 11 mm - main body lengths contralateral limb 74mm ipsilateral limb 133 mm - delivery system 18 FR; Catalog no. TFB-24-74.
- Product code
- MIH System, Endovascular Graft, Aortic Aneurysm Treatment
- Reason for recall
- Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use.
- FDA root cause
- Other
- Action
- The firm issued recall letters to customers on 6/10/04. The recall was extended and URGENT EXTENSION OF PRODUCT RECALL notices were sent out on June 23, 2004.
- Quantity in commerce
- 18
- Distribution
- Nationwide, Canada, Denmark and Japan.
- Product codes and lots
- Lots 1352808, 1357063, F1537398, F1538637, F1539268, F1541241, F1549741, F1543305, F1545887, F1553587, F1560030, F1560031, F1565092, F1567916, F1565091, F1570152, 1357066 and 1357065.