Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1194-04

Cook brand Zenith AAA Iliac leg graft with the H&L-B One-Shot Introducer System, preloaded iliac leg graft...

Cook Incorporated / initiated 2004-06-10 / Terminated

Cook, Inc. began a recall of Cook brand Zenith AAA Iliac leg graft with the H&L-B One-Shot Introducer System, preloaded iliac leg graft - distal cuff diameter 16 mm - usable length 71mm -... on . The reason FDA records is "Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1194-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Cook brand Zenith AAA Iliac leg graft with the H&L-B One-Shot Introducer System, preloaded iliac leg graft - distal cuff diameter 16 mm - usable length 71mm - delivery system 14 FR; Catalog no. TFLE-16-71.
Product code
MIH System, Endovascular Graft, Aortic Aneurysm Treatment
Reason for recall
Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use.
FDA root cause
Other
Action
The firm issued recall letters to customers on 6/10/04. The recall was extended and URGENT EXTENSION OF PRODUCT RECALL notices were sent out on June 23, 2004.
Quantity in commerce
96
Distribution
Nationwide, Canada, Denmark and Japan.
Product codes and lots
Lots 1366580, 1366579, 1366582, 1360764, 1360766, 1366578, 1360760, F1558983, F1558984, F1558986, F1556851, F1556852, F1559095, F1559096, F1559097, F1559098, F1558985, F1558987, F1564117, F1564118, F1564119, F1564183, F1564376, F1565664, F1565665, F1565666, F1565667, F1565668, F1565669, F1565670, F1565671, F1564179, F1564180, F1564181, F1564182, F1564184, F1564185, F1564186, F1564187, F1564188, F1564189, F1564377, F1564378, F1564379, F1564380, F1565778, F1565779, F1565780, F1565781, F1565782, F1570176, F1570177, F1569280, F1569281, F1569282, F1569283, F1571046, F1571047, F1571048, F1571049, F1571050, F1571053, F1569251, F1569252, F1569253, F1569254, F1572993, F1572994, F1572995, F1572996, F1572997, F1572998, F1572999, F1575325, F1575326, F1575327, F1575328, F1575195, F1575196, F1575197, F1575198, F1575199, F1569311, F1573000, F1573001, F1575324, 1360758, 1360761, 1360762, 1360759, 1360763, 1366575, 1366577, 1366584, 1360754 and 1360767.