Z-1118-04

Nichols Advantage 25-Hydroxy Vitamin D Assay, Catalog number 62-7033

Nichols Institute Diagnostics / initiated 2004-06-30 / Terminated

Nichols Institute Diagnostics began a recall of Nichols Advantage 25-Hydroxy Vitamin D Assay, Catalog number 62-7033 on . The reason FDA records is "Lower than expected results are obtained". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1118-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Nichols Institute Diagnostics
Product
Nichols Advantage 25-Hydroxy Vitamin D Assay, Catalog number 62-7033.
Product code
MRG System, Test, Vitamin D
Reason for recall
Lower than expected results are obtained.
FDA root cause
Other
Action
Firm sent Customer Bulletin No 04-09-AD on June 30, 2004. It addresses the issues and advises customers as to alternative actions, including appending results with a caution statement regarding low recovery.
Quantity in commerce
Not Applicable.
Distribution
Nationwide. Canada, Australia, Greece, Denmark, Portugal, Sweden, Brazil, Italy, japan, Spain, Germany, France.
Product codes and lots
Not limited to specific lots.