Z-1159-04
Alsius CoolGuard 3000 Temperature Regulation System, Model CoolGuard 3000
Alsius Corp. / initiated 2004-06-21 / Terminated
Alsius Corporation began a recall of Alsius CoolGuard 3000 Temperature Regulation System, Model CoolGuard 3000 on . The reason FDA records is "Higher than expected rate of pump alarms occur under specific conditions". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1159-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alsius Corp.
- Product
- Alsius CoolGuard 3000 Temperature Regulation System, Model CoolGuard 3000
- Product code
- NCX System, Hypothermia, Intravenous, Cooling
- Reason for recall
- Higher than expected rate of pump alarms occur under specific conditions.
- FDA root cause
- Other
- Action
- Firm replaced circuit board having a new resistor. Recall is complete.
- Quantity in commerce
- 15
- Distribution
- MA, PA, VA, FL, AL, MO, TX
- Product codes and lots
- All codes.