Z-1303-04

Roche brand Reflotron GPT (ALT) 30 test strips; catalog # 745138

Roche Diagnostics GmbH / initiated 2004-07-14 / Terminated

Roche Diagnostics Corp. began a recall of Roche brand Reflotron GPT (ALT) 30 test strips; catalog # 745138 on . The reason FDA records is "Coating defect may result is falsely decreased patient and/or control results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1303-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche brand Reflotron GPT (ALT) 30 test strips; catalog # 745138.
Product code
CIT Nadh Oxidation/Nad Reduction, Ast/Sgot
Reason for recall
Coating defect may result is falsely decreased patient and/or control results.
FDA root cause
Other
Action
Recall letters dated 7/14/04 were sent to each customer asking that use of these lots be stopped immediately. Testing labs were asked to notify the attending physician for any patients tested with these test strips.
Quantity in commerce
197 boxes
Distribution
United States.
Product codes and lots
Lot 23717132; exp 5/31/2005.