Z-1329-04

Protege GPS Biliary System (6Fr 10mm x 30mm) - Product number SER6-10-30-135 (.018) (Biliary)

Medtronic, Inc. / initiated 2004-07-09 / Terminated / root cause: labeling and UDI

Ev3 began a recall of Protege GPS Biliary System (6Fr 10mm x 30mm) - Product number SER6-10-30-135 (.018) (Biliary) on . The reason FDA records is "Ev3 had received customer complaints indicating that the stent device could not be used with an .035' guidewire. An investigation found that 6Fr, 135 cm (.018') stent system were labeled as a 120 cm (.035') stent system...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1329-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Protege GPS Biliary System (6Fr 10mm x 30mm) - Product number SER6-10-30-135 (.018) (Biliary)
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
Ev3 had received customer complaints indicating that the stent device could not be used with an .035' guidewire. An investigation found that 6Fr, 135 cm (.018') stent system were labeled as a 120 cm (.035') stent system and that 6Fr 120 cm (.035') were labeled as 6Fr, 135 cm (.018').
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Send notification letter to direct accounts and an EV3 representative will contact the direct account within the next few days to arrrange for the return of all unused product.
Quantity in commerce
10
Distribution
Eleven (11) devices were distributed to seven (7) U.S. customers. Ten (10) devices were distributed to seven (7) clinical study sites in the U.S.
Product codes and lots
Lot # 672655