Z-1338-04

Accura System for Blood Filtration, product codes 5M5660 and ACCURA01 (international); a hemofiltration...

Baxter Healthcare Renal Div / initiated 2004-07-26 / Terminated

Baxter Healthcare Renal Div began a recall of Accura System for Blood Filtration, product codes 5M5660 and ACCURA01 (international); a hemofiltration system; Baxter Healthcare Corporation, Renal Division... on . The reason FDA records is "The Accura machine display screen may go blank during patient treatment. There is a potential risk to the patient in that the operator may try to manipulate the controls in response to a blank screen, which could affect...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1338-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Renal Div
Product
Accura System for Blood Filtration, product codes 5M5660 and ACCURA01 (international); a hemofiltration system; Baxter Healthcare Corporation, Renal Division, McGaw Park, IL 60085
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
The Accura machine display screen may go blank during patient treatment. There is a potential risk to the patient in that the operator may try to manipulate the controls in response to a blank screen, which could affect the current patient's treatment
FDA root cause
Other
Action
Baxter sent an Urgent Device Correction letter dated 7/26/04 to the direct accounts on the same date, to the attention of the Hemodialysis Unit and Hospital Administrator. The letters informed the accounts of the potential problem and provided instructions on the procedure to follow in the event the screen goes blank during patient use. Any questions were directed to Baxter Instrument Services at 1-800-553-6898, select prompt 3, option 4.
Quantity in commerce
85 units
Distribution
Nationwide and internationally to Belgium, China and Hong Kong
Product codes and lots
product code 5M5660, serial numbers 551-556, 569, 570, 582-587, 600-604, 623, 628, 649, 650, 656-658, 667-670, 677, 685-688, 695-697, 709-712, 719-722, 729, 730, 732, 740, 741  Product code ACCURA01: serial numbers 519, 520, 532-534, 536-541, 550, 557-560, 571-574, 579-581, 588, 589, 599, 617, 659-661, 676, 678, 679