Protege GPS Biliary System (6Fr 10mm x 30mm) - Product number SERB65-10-30-120 (.035) (Biliary)
Ev3 began a recall of Protege GPS Biliary System (6Fr 10mm x 30mm) - Product number SERB65-10-30-120 (.035) (Biliary) on . The reason FDA records is "Ev3 had received customer complaints indicating that the stent device could not be used with an .035' guidewire. An investigation found that 6Fr, 135 cm (.018') stent system were labeled as a 120 cm (.035') stent system...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1330-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Protege GPS Biliary System (6Fr 10mm x 30mm) - Product number SERB65-10-30-120 (.035) (Biliary)
- Product code
- FGE Stents, Drains And Dilators For The Biliary Ducts
- Reason for recall
- Ev3 had received customer complaints indicating that the stent device could not be used with an .035' guidewire. An investigation found that 6Fr, 135 cm (.018') stent system were labeled as a 120 cm (.035') stent system and that 6Fr 120 cm (.035') were labeled as 6Fr, 135 cm (.018').
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Send notification letter to direct accounts and an EV3 representative will contact the direct account within the next few days to arrrange for the return of all unused product.
- Quantity in commerce
- 11
- Distribution
- Eleven (11) devices were distributed to seven (7) U.S. customers. Ten (10) devices were distributed to seven (7) clinical study sites in the U.S.
- Product codes and lots
- Lot numbers: 257984, 257985, and 257986.