Z-1430-04
Roche/Hitachi Modular Analytical D Module DAT clinical chemistry analyzer; catalog number 04429338160
Roche Diagnostics GmbH / initiated 2004-08-06 / Terminated / root cause: software
Roche Diagnostics Corp. began a recall of Roche/Hitachi Modular Analytical D Module DAT clinical chemistry analyzer; catalog number 04429338160 on . The reason FDA records is "A software defect will allow the reporting out of believable, but clinically significant, erroneous results under certain conditions". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1430-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche/Hitachi Modular Analytical D Module DAT clinical chemistry analyzer; catalog number 04429338160.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- A software defect will allow the reporting out of believable, but clinically significant, erroneous results under certain conditions.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Recall letters dated 8/6/04 were sent to each customer advising them of the problem and giving instructions on how to avoid it.
- Distribution
- United States.
- Product codes and lots
- All units.