Z-1405-04
Guardian Distal Femur Axial Pin
Wright Medical Technology Inc / initiated 2004-07-14 / Terminated
Wright Medical Technology Inc began a recall of Guardian Distal Femur Axial Pin on . The reason FDA records is "Potential for self -locking axial pin to dislodge and 'back out'". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1405-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Wright Medical Technology Inc
- Product
- Guardian Distal Femur Axial Pin
- Product code
- JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
- Reason for recall
- Potential for self -locking axial pin to dislodge and 'back out'.
- FDA root cause
- Other
- Action
- The firm sent out product correction notifications to implanting surgeons by letter dated 07/14/2004 to notify the surgeons of the dislodgement issue.
- Quantity in commerce
- 44
- Distribution
- Nationwide. Ontario, and Taiwan, ROC
- Product codes and lots
- Part No. 25002111, Serial No. 02217952 Part No. 25002112, Serial No. 01214391 Part No. 25002112, Serial No. 01215073 Part No. 25002113, Serial No. 01214533 Part No. 25002114, Serial No. 01215034 Part No. 25002114, Serial No. 01215074