Z-1376-04

ProSpecT Cryptosporidium Microplate Assay. Catalog #2454096

Remel Inc / initiated 2002-05-09 / Terminated

Remel, Inc. began a recall of ProSpecT Cryptosporidium Microplate Assay. Catalog #2454096 on . The reason FDA records is "A potential stability problem has been identified with the Positive Controls in the referenced kit lots of the ProSpecT Cryptosporidium Microplate Assay. The Positive Control may NOT perform as intended through its...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1376-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Remel Inc
Product
ProSpecT Cryptosporidium Microplate Assay. Catalog #2454096.
Product code
MHJ Cryptosporidium Spp.
Reason for recall
A potential stability problem has been identified with the Positive Controls in the referenced kit lots of the ProSpecT Cryptosporidium Microplate Assay. The Positive Control may NOT perform as intended through its expiration date and may invalidate an otherwise acceptable run. Quality control testing indicates the other reagents in these kits are performing as intended.
FDA root cause
Other
Action
A potential stability problem has been identified with the positive control in the ProSpecT CryptoSporidium Microplate Assay. The positive control may not perform as intended through its expiration date and may invalidate an otherwise acceptable run.
Quantity in commerce
388 kits (187 - 910717, 201 - 912101)
Distribution
in the US
Product codes and lots
910717 912101