General Electric Med Systems began a recall of SMV FX-80 Nuclear Camera on . The reason FDA records is "The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector's radial drive assembly can back out of the bearing block that contains...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1394-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- SMV FX-80 Nuclear Camera
- Product code
- IYX Camera, Scintillation (Gamma)
- Reason for recall
- The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector's radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become loose, fail, or back out.
- FDA root cause
- Other
- Action
- The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector''s radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become lose, fail, or back out.
- Quantity in commerce
- 41
- Distribution
- Inside the US. Manufactured and distributed from 1993 - 2000.
- Product codes and lots
- 00000000000172, 00000000000173, 00000000000188, 00000000000203, 00000000000209, 00000000000214, 00000000000223, 0070090080-103, 0070090080-106, 0070090080-107, 0070090080-112, 0070090080-115, 0070090080-116, 0070090080-123, 0070090080-124, 0070090080-125, 0070090080-126, 0070090080-127, 0070090080-138, 0070090080-141, 0070090080-146, 0070090080-150, 0070090080-152, 0070090080-158, 0070090080-161, 0070090080-162, 0070090080-163, 0070090080-164, 0070090080-165, 0070090080-167, 0070090080-168, 0070090080-170, 0070090080-172, 0070090080-173, 0070090080-179, 0070090080-180, 0070090080-188, 0070090080-203, 0070090080-209, 0070090080-96, 0070090080-P298