Ref. No. 523770 Hemoclip, 15 Large Titanium Clips/Cartridge, and Ref. No. 523870 Hemoclip 10 Large Titanium...

Teleflex Medical / initiated 2003-10-24 / Terminated / root cause: packaging

Weck began a recall of Ref. No. 523770 Hemoclip, 15 Large Titanium Clips/Cartridge, and Ref. No. 523870 Hemoclip 10 Large Titanium Clips/Cartridge on . The reason FDA records is "Hole in the sterile unit blister pack that could compromise sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1387-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Teleflex Medical
Product
Ref. No. 523770 Hemoclip, 15 Large Titanium Clips/Cartridge, and Ref. No. 523870 Hemoclip 10 Large Titanium Clips/Cartridge
Product code
HBT Applier, Hemostatic Clip
Reason for recall
Hole in the sterile unit blister pack that could compromise sterility.
Root cause tag
packaging
FDA root cause
Other
Action
The consignees were notified of the recall via letter beginning October 24, 2003 asking for return of the product.
Quantity in commerce
91,869 units for all products under recall
Distribution
Nationwide and to accounts in Canada, Europe Asia, and South/Latin America
Product codes and lots
All lots manufactured between 2001-8 through 2002-12 for the 15 and 10 Large Titanium Clips and additional lots 799983, 816529, 826816, and 840190 for the 10 Large Titanium Clips