Z-1447-04
Zimmer brand NEXGEN Complete Knee Solution Rotating Hinge Knee Tibial Component, precoat, nonmodular, size 2...
Zimmer GmbH / initiated 2004-08-03 / Terminated
Zimmer Inc. began a recall of Zimmer brand NEXGEN Complete Knee Solution Rotating Hinge Knee Tibial Component, precoat, nonmodular, size 2, sterile, for cemented use only; Cat. No... on . The reason FDA records is "The polyethylene tibial bushing was omitted from the taper of the tibial plates". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1447-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer brand NEXGEN Complete Knee Solution Rotating Hinge Knee Tibial Component, precoat, nonmodular, size 2, sterile, for cemented use only; Cat. No. 5880-02-02.
- Product code
- KRO Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
- Reason for recall
- The polyethylene tibial bushing was omitted from the taper of the tibial plates.
- FDA root cause
- Other
- Action
- Consignees were notified via recall letter dated August 3, 2004 and sent via certified mail, return receipt requested. A dear doctor letter dated August 25, 2004 was sent to the implanting physicians.
- Distribution
- Nationwide (California, Florida, Illinois, Indiana, Louisiana, Michigan, Nevada, New Jersey, New York, North Carolina, Ohio, Texas, Wisconsin), and Germany.
- Product codes and lots
- Lot 60037871; Exp. June 2008.