Z-1473-04
VITROS CRP Slides: (1) Cat # 192 6740, 250 slides per box, and (2) Cat # 809 7990, 90 slides per box. Firm on...
Ortho-Clinical Diagnostics, Inc. / initiated 2004-07-28 / Terminated / root cause: software
Ortho-Clinical Diagnostics began a recall of VITROS CRP Slides: (1) Cat # 192 6740, 250 slides per box, and (2) Cat # 809 7990, 90 slides per box. Firm on the label: Ortho-Clinical Diagnostics, Inc... on . The reason FDA records is "The Immuno-rate (IR) wash detection algorithm may result in inappropriate wash detections". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1473-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS CRP Slides: (1) Cat # 192 6740, 250 slides per box, and (2) Cat # 809 7990, 90 slides per box. Firm on the label: Ortho-Clinical Diagnostics, Inc., Rochester, NY 14626. The CRP Slides are for use on the Vitros 250/250AT and/or Vitros 950/950AT Clinical Chemistry Systems. For in vitro diagnostic use. Vitros CRP slides quantitatively measure C-reactive protein concentration in serum and plasma.
- Product code
- DCK C-Reactive Protein, Antigen, Antiserum, And Control
- Reason for recall
- The Immuno-rate (IR) wash detection algorithm may result in inappropriate wash detections.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Letters dated 7/28/2004. Distributors instructed to destroy product. Users instructed to use revised IR wash tolerance values (provided with letter). A Calibration Diskette, Data Release Version 5384 will also be provided to end users.
- Quantity in commerce
- Cat 192 6740 - 5,912 cartons x 250 slides; Cat #809 7990 - 10,344 x 90 slides
- Distribution
- DOMESTIC distribution (nationwide): Product shipped to 6 distributors, 140 hospitals, and 6 U.S. government accounts. FOREIGN distribution: Product shipped to 22 Ortho-Clinical/J&J affiliates in Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Peru, Ecuador, India, Japan, Mexico, Puerto Rico, Panama, Singapore, Venezuela, England, France, Germany, Italy and Spain. The foreign affiliates further distributed product.
- Product codes and lots
- Coatings 401 thru 409 (Generations 92, 93, 94, 99, 01, 02, 03, 04, and 05)