Z-1434-04

Roche/Hitachi ALT (ALAT/GPT) Reagent Kit for use on the Roche/Hitachi MODULAR ANALYTICS P Module; Catalog...

Roche Diagnostics GmbH / initiated 2004-08-20 / Terminated

Roche Diagnostics Corp. began a recall of Roche/Hitachi ALT (ALAT/GPT) Reagent Kit for use on the Roche/Hitachi MODULAR ANALYTICS P Module; Catalog number 1877470 (11877470216) on . The reason FDA records is "Rotation of the instrument reagent compartment can cause a bubble to form in the neck of the reagent bottle causing premature liquid level detection and short sampling of the reagent, resulting in falsely low or zero...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1434-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche/Hitachi ALT (ALAT/GPT) Reagent Kit for use on the Roche/Hitachi MODULAR ANALYTICS P Module; Catalog number 1877470 (11877470216).
Product code
CKH Acid Phosphatase, Beta Glycerophosphate
Reason for recall
Rotation of the instrument reagent compartment can cause a bubble to form in the neck of the reagent bottle causing premature liquid level detection and short sampling of the reagent, resulting in falsely low or zero patient results.
FDA root cause
Other
Action
Consignees were mailed recall notices dated 8/20/04 and were instructed to either reduce the initial reagent bottle volume prior to use, or to request replacement product.
Quantity in commerce
728 non-expired kits.
Distribution
Nationwide.
Product codes and lots
All lots expiring on or before 11/05.