Roche/Hitachi AST (ASAT/GOT) Reagent Kit for use on the Roche/Hitachi MODULAR ANALYTICS P Module; Catalog...
Roche Diagnostics Corp. began a recall of Roche/Hitachi AST (ASAT/GOT) Reagent Kit for use on the Roche/Hitachi MODULAR ANALYTICS P Module; Catalog number 1928511 (11928511216) on . The reason FDA records is "Rotation of the instrument reagent compartment can cause a bubble to form in the neck of the reagent bottle causing premature liquid level detection and short sampling of the reagent, resulting in falsely low or zero...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1435-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche/Hitachi AST (ASAT/GOT) Reagent Kit for use on the Roche/Hitachi MODULAR ANALYTICS P Module; Catalog number 1928511 (11928511216).
- Product code
- CIT Nadh Oxidation/Nad Reduction, Ast/Sgot
- Reason for recall
- Rotation of the instrument reagent compartment can cause a bubble to form in the neck of the reagent bottle causing premature liquid level detection and short sampling of the reagent, resulting in falsely low or zero patient results.
- FDA root cause
- Other
- Action
- Consignees were mailed recall notices dated 8/20/04 and were instructed to either reduce the initial reagent bottle volume prior to use, or to request replacement product.
- Quantity in commerce
- 757non-expired kits.
- Distribution
- Nationwide.
- Product codes and lots
- All lots expiring on or before 7/05.