ARCHITECT B12 Reagents, List 6C09-20 (4 x 100 tests) and List 6C09-25 (100 tests); Abbott Laboratories...
Abbott Laboratories HPD/ADD/GPRD began a recall of ARCHITECT B12 Reagents, List 6C09-20 (4 x 100 tests) and List 6C09-25 (100 tests); Abbott Laboratories, Abbott Park, IL 60064 USA; Abbott, Max-Planck-Ring 2... on . The reason FDA records is "Changes in the ARCHITECT B12 Reagent lots may adversely affect the shape of the calibration curve, over time. These changes may produce controls out of range low and similar decreases in patient sample values prior to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1452-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ARCHITECT B12 Reagents, List 6C09-20 (4 x 100 tests) and List 6C09-25 (100 tests); Abbott Laboratories, Abbott Park, IL 60064 USA; Abbott, Max-Planck-Ring 2, 65205 Wiesbaden, Germany
- Product code
- LIG Radioassay, Intrinsic Factor Blocking Antibody
- Reason for recall
- Changes in the ARCHITECT B12 Reagent lots may adversely affect the shape of the calibration curve, over time. These changes may produce controls out of range low and similar decreases in patient sample values prior to current expiration of the reagents.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Abbott affiliates in Canada, Germany, Singapore, Hong Kong, Australia, New Zealand and Japan were e-mailed copies of the recall letter on 8/19/04 for translation where needed and dissemination to their accounts receiving the suspect lots. The end users were advised of the potential for changes in the reagents over time that affect the shape of the calibration curve, producing controls out of range low and similar decreases in patient sample values. The accounts were instructed to manually update the expiration dates for the reagent lots according to the chart given in the recall letter for any of the affected lots remaining in inventory. The expiration date reported by the Architect instrument must also be manually updated.
- Quantity in commerce
- 5754 kits
- Distribution
- Canada, Germany, Singapore, Hong Kong, Australia, New Zealand and Japan.
- Product codes and lots
- List 6C09-20, lots 11002M300, 11003M300, 16051M200, 18288M300; List 6C09-25, lots 11002M301, 16051M201, 18288M301