Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1474-04

XSYSTEMS Dilution Buffer, list number 9519-02; for In Vitro diagnostic use, Bovine gamma globulin in...

Abbott Medical / initiated 2004-09-01 / Terminated

Abbott Laboratories HPD/ADD/GPRD began a recall of XSYSTEMS Dilution Buffer, list number 9519-02; for In Vitro diagnostic use, Bovine gamma globulin in Phosphate buffer, with Sodium Azide as a preservative; 950... on . The reason FDA records is "The XSYSTEMS Dilution Buffer when used with TDx/TDxFLx Benzodiazepines assay can cause shifts in control values and patient results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1474-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
XSYSTEMS Dilution Buffer, list number 9519-02; for In Vitro diagnostic use, Bovine gamma globulin in Phosphate buffer, with Sodium Azide as a preservative; 950 mL bottle, 4 bottles per carton; Abbott Laboratories, Abbott Park, IL 60064 USA
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The XSYSTEMS Dilution Buffer when used with TDx/TDxFLx Benzodiazepines assay can cause shifts in control values and patient results.
FDA root cause
Other
Action
Abbott sent Device Correction letters dated 9/1/04 to all customers receiving the TDx/TDxFLx Benzodiazepines assay, list 9674, informing them of the shift in control values and patient results when used with the XSYSTEMS Dilution Buffer, list 9519-02. The letter gave examples of the effects observed in the benzodiazepines assay performance, and listed the following actions to be taken until Abbott completes their investigation: * During the use of a single lot of buffer, recalibrate the TDx/TDxFLx Benzodiazepines assay every 14 days. * Perform a new calibration run on the TDx/TDxFLx Benzodiazepines assay when a new lot number of XSYSTEMS Dilution Buffer is used. * These actions are in addition to the package insert calibration and quality control requirements. * Follow your laboratory procedures for notification of the health care providers that you serve. Any questions were directed to Abbott''s Customer Service at 1-877-4ABBOTT.
Quantity in commerce
22,442 cartons
Distribution
Nationwide and internationally to Argentina, Australia, Brazil, Chile, Colombia, Costa Rica, El Salvador, Germany, Guatemala, Honduras, Hong Kong, Mexico, Panama, Peru, Puerto Rico, Singapore, South Korea, Taiwan, Thailand, the United Kingdom, Uruguay and Venezuela
Product codes and lots
list number 9519-02, all lots