Z-1499-04

Neonatal Total Galactose Test Kits. Reagents for 960 (4800) assays, Item #NG-4000

Revvity Health Sciences, Inc. / initiated 2004-09-08 / Terminated / root cause: manufacturing process

PerkinElmer LAS Inc began a recall of Neonatal Total Galactose Test Kits. Reagents for 960 (4800) assays, Item #NG-4000 on . The reason FDA records is "There is an increase in the number of normal specimen results below the lowest calibrator and based on a report that the normal controls were running low, sometimes out of the specified range". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1499-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Revvity Health Sciences, Inc.
Product
Neonatal Total Galactose Test Kits. Reagents for 960 (4800) assays, Item #NG-4000.
Product code
JIA Enzymatic Methods, Galactose
Reason for recall
There is an increase in the number of normal specimen results below the lowest calibrator and based on a report that the normal controls were running low, sometimes out of the specified range.
Root cause tag
manufacturing process
FDA root cause
Other
Action
The firm contacted their consignees by letter sent 9/8/04.
Quantity in commerce
49 kits
Distribution
The kits were distributed to 5 consignees, located in Indianapolis, IN; Jefferson City, MO; Austin, TX; Madison, WI; and Minneapolis, MN.
Product codes and lots
Lot #244468