Emission Computed Tomography System. Including Model numbers: 2200967, ASM001080, ASM001083, ASM001092...

GE Healthcare / initiated 2004-09-03 / Terminated

General Electric Med Systems LLC began a recall of Emission Computed Tomography System. Including Model numbers: 2200967, ASM001080, ASM001083, ASM001092, ASM001099, ASM000085, 10036800109, and Varicam on . The reason FDA records is "GE Healthcare received reports of intermittent low ejection fraction (EF) results for MUGA (multi-gated acquisitions) studies on Millennium VG systems. This issue can occasionally lead to distortions in the volume curve...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0012-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Emission Computed Tomography System. Including Model numbers: 2200967, ASM001080, ASM001083, ASM001092, ASM001099, ASM000085, 10036800109, and Varicam.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
GE Healthcare received reports of intermittent low ejection fraction (EF) results for MUGA (multi-gated acquisitions) studies on Millennium VG systems. This issue can occasionally lead to distortions in the volume curve in some MUGA studies, and result in calculated EF values lower (never higher) than actual values. A low EF value could lead to a physician decision to stop needed treatment.
FDA root cause
Other
Action
A letter dated August 5, 2004 was released to customers describing the issue and recommending that users pay special attention to any volume curve which appears to be abnormal or any EF values that appear to be lower than expected, and not to rely soley on this value for diagnosis purposes until they recieve the SW patch. A SW patch will be distributed shortly to all affected systems.
Quantity in commerce
853 units
Distribution
Nationwide throughout the United States and worldwide.
Product codes and lots
Each unit has a unique serial code. All serial numbers are being recalled.