RESOLVE Hemoglobin Test Kits packed and distributed under the Perkin Elmer Life and Analytical Sciences...
PerkinElmer LAS Inc began a recall of RESOLVE Hemoglobin Test Kits packed and distributed under the Perkin Elmer Life and Analytical Sciences label, Kit Codes: FR-9120 - 120 tests per kit; FR-9400... on . The reason FDA records is "The product insert for the test kits is incorrect with regard to the acceptable storage temperature for the Agarose IEF gel component. The kit and IEF gel labels indicate the correct storage temperature of 2-8 degrees...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0047-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Revvity Health Sciences, Inc.
- Product
- RESOLVE Hemoglobin Test Kits packed and distributed under the Perkin Elmer Life and Analytical Sciences label, Kit Codes: FR-9120 - 120 tests per kit; FR-9400 - 360 tests per kit; and FR-9360 - 3600 tests per kit.
- Product code
- GKA Abnormal Hemoglobin Quantitation
- Reason for recall
- The product insert for the test kits is incorrect with regard to the acceptable storage temperature for the Agarose IEF gel component. The kit and IEF gel labels indicate the correct storage temperature of 2-8 degrees C. The product insert incorrectly indicates that the product may be stored at 2-30 degrees C.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- The firm contacted their consignees by letter dated 9/14/04.
- Quantity in commerce
- All lots of test kits manufactured by the firm which have an active shelf life are subject to recall.
- Distribution
- Nationwide (The product was distributed to firms located in AR, AZ, CA, CO,FL, GA, HI, IA, IL, IN, KS, KY,LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV )and worldwide (to 12 foreign accounts).
- Product codes and lots
- FR-9120 - Lot Number: 243309 FR-9360 - Lot Numbers: 213926, 213932, 213939, 213944, 220789, 220794, 220804, 220809, 220799, 220814, 220819, 220824, 220829, 220834, 220839, 220848, 220855, 220860, 220865, 220870, 220875, 220882, 220892, 220898, 228304, 228309, 228314, 228319, 228325, 228330, 228335, 228341, 228348, 228353, 228358, 228363, 228368, 228374, 228379, 228386, 228391, 228396, 229401, 228406, 234161, 234172, 234177, 234182, 234190, 234195, 234200, 234205, 234210, 234222, 234231, 234240, 234247, 234252, 234257, 234262, 238917, 234267, 234272, 234277, 238920, 238923, 238926, 238929, 238932, 238946, 238950, 238953, 238960, 238963, 242331, 242334, 242334. FR-9400 - Lot Numbers: 213941, 220791, 220901, 220806, 220811, 220801, 220816, 220826, 220831, 220836, 220841, 220857, 220862, 220902, 220872, 220894, 220900, 228306, 228316, 228327, 228332, 228344, 228350, 228355, 228360, 228365, 228376, 228381, 228388, 228398, 228403, 228408, 2234192.