Z-1494-04
Stockinette, Double Ply, Econo Weave, 6 X 54'', 6 X 60'', 6 X 72'', 8 X 48'', 8 X 72'', Ref #9984-54...
Deroyal Surgical -Rose Hill / initiated 2004-09-09 / Terminated / root cause: packaging
Deroyal Surgical -Rose Hill began a recall of Stockinette, Double Ply, Econo Weave, 6 X 54'', 6 X 60'', 6 X 72'', 8 X 48'', 8 X 72'', Ref #9984-54, #9984-54, #9985-60, #9986-72, #9987-48, #9988-60, #9988-72 on . The reason FDA records is "Surgical drapes and covers were packaged in containers with incomplete seals which compromise product sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1494-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Deroyal Surgical -Rose Hill
- Product
- Stockinette, Double Ply, Econo Weave, 6 X 54'', 6 X 60'', 6 X 72'', 8 X 48'', 8 X 72'', Ref #9984-54, #9984-54, #9985-60, #9986-72, #9987-48, #9988-60, #9988-72
- Product code
- KKX Drape, Surgical, Antimicrobial
- Reason for recall
- Surgical drapes and covers were packaged in containers with incomplete seals which compromise product sterility.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- The recalling firm ceased distribution of Stockinette products on 8/26/04 and subsequently notified distributors and direct accounts by letter 9/9/04. The notification was product and lot specific and advised consignees of compromised sterility. Consignees were advised to return response form and product for credit. Distributors were asked to forward the recall notice and respond with a consignee list if they had not notified their accounts. Sub accounts and end users were notified by letter 9/20/04.
- Quantity in commerce
- 395 cases
- Distribution
- Product was sold to distributors and direct accounts nationwide and to distributors in Ireland, Peru and Israel.
- Product codes and lots
- Lots: 9984-54 9985-60 9991-60 9986-72 9987-48 9988-72