Z-0011-05

Medtronic MiniMed Paradigm Model 512

Medtronic, Inc. / initiated 2004-09-14 / Terminated

Medtronic MiniMed began a recall of Medtronic MiniMed Paradigm Model 512 on . The reason FDA records is "Several complaints of pumps exhibiting A47 alarms were received. Investigation revealed the alarms occurred when the pump attempted to display text that exceeded the maximum display length of the pump screen. The...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0011-05
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Medtronic MiniMed Paradigm Model 512
Product code
LZG Pump, Infusion, Insulin
Reason for recall
Several complaints of pumps exhibiting A47 alarms were received. Investigation revealed the alarms occurred when the pump attempted to display text that exceeded the maximum display length of the pump screen. The investigation revealed this only occurs when the pump is programmed to display information in Spanish.
FDA root cause
Other
Distribution
Nationwide & Spain
Product codes and lots
Paradigm model 512pumps with serial numbers that end in '-Axx4' . Where 'x' can be any number or letter.