Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0052-05

BD Vactuainer SST glass and plastic tubes

Becton Dickinson & Co / initiated 2004-08-20 / Terminated

Becton Dickinson & Company began a recall of BD Vactuainer SST glass and plastic tubes on . The reason FDA records is ""Please be aware tht this not a new recall. The firm has taken action; but, due to administrative issues additional products are now being classified by the agency". There is an apparent bias in the determination of...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0052-05
Status
Terminated
Initiated
Posted
Recalling firm
Becton Dickinson & Co
Product
BD Vactuainer SST glass and plastic tubes.
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
"Please be aware tht this not a new recall. The firm has taken action; but, due to administrative issues additional products are now being classified by the agency". There is an apparent bias in the determination of Total T3, Total T4, HBsAg, Folate BA, VB 12, BR, FSH and Cortisol hormones used to determine thyroid disease states, and adrenal gland function. The immunoassay results are high when
FDA root cause
Other
Action
The firm sent out a Technical Bulletin to all consignees of the specific instrument companies involved. Mailing initial were done 9/17/2004 and a revised mailing was sent 9/23/2004. Global communication will commense 10/5/2004.
Quantity in commerce
There is no linkage to any particular lot of BD product
Distribution
Worldwide distribution.
Product codes and lots
There is no linkage to any particular lot of BD product.