Z-0037-05
NOVAPLUS Electrosurgical Grounding Pads, Split with Cord Catalog No. 7179V
3M Company / initiated 2004-09-30 / Terminated
3M Company / Medical Division began a recall of NOVAPLUS Electrosurgical Grounding Pads, Split with Cord Catalog No. 7179V on . The reason FDA records is "Improper wire placement from the patient plate to the ESU connector plug results in an intermittent displacement connection". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0037-05
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- 3M Company
- Product
- NOVAPLUS Electrosurgical Grounding Pads, Split with Cord Catalog No. 7179V
- Product code
- JOS Electrode, Electrosurgical
- Reason for recall
- Improper wire placement from the patient plate to the ESU connector plug results in an intermittent displacement connection.
- FDA root cause
- Other
- Action
- A recall notification letter dated 09/30/2004 was sent to Hospitals and Health Care Distributors instructing consignees to immedately remove all effective product from lot 2007-06 HP.
- Quantity in commerce
- 384 cases containing 40 pads each for a total of 15,360 pads
- Distribution
- The subject product lot was distributed to the following key 3M distributors: Owens & Minor, Cardinal Health, McKesson, Professional Health Supply, and the Burrows Company.
- Product codes and lots
- Lot 2007-06 HP